Abstract

Radiographic assessment of lower-limb alignment (LLA) is important for predicting joint health and surgical outcomes in total knee arthroplasty. Traditional measurement methods are manual and time-consuming, while recent machine learning approaches typically rely on locating a fixed set of anatomical landmarks. This dependence limits flexibility and may require re-annotation when clinical definitions change. To address this, we propose an automated workflow using Implicit Neural Shape Functions (INSF). Rather than relying on explicit landmark coordinates, we encode the anatomy into a compact latent space and regress clinical alignment measurements directly from these latent codes. This architecture allows for rapid extendability to new tasks without altering the backbone representation. We trained our method on an internal dataset of 566 knee radiographs, each annotated with the outline of the femur and tibia. We evaluated it on both an internal test dataset of 50 patients and a separate external set of 402 preoperative cases from the MRKR dataset. Performance was comparable to state-of-the-art landmark-based methods and manual agreement, while offering a flexible shape representation that can be extended to additional measurement tasks. The MRKR measurement data is available at https://huggingface.co/datasets/imedslab/mrkr-knee-alignment.

Links to Paper and Supplementary Materials

Main Paper (Open Access Version): https://papers.miccai.org/miccai-2026/paper/5630_paper.pdf

SharedIt Link: Not yet available

SpringerLink (DOI): Not yet available

Supplementary Material: Not Submitted

Link to the Code Repository

N/A

Link to the Dataset(s)

https://huggingface.co/datasets/imedslab/mrkr-knee-alignment

BibTex

@InProceedings{HuZhi_LandmarkFree_MICCAI2026,
        author = { Hu, Zhisen AND Kemppainen, Antti AND Johnson, David AND Panfilov, Egor AND Nguyen, Huy Hoang AND Cootes, Timothy AND Lindner, Claudia AND Tiulpin, Aleksei},
        title = { { Landmark-Free Assessment of Lower-Limb Alignment with Implicit Neural Shape Functions from Knee Radiographs } },
        booktitle = {Medical Image Computing and Computer Assisted Intervention -- MICCAI 2026},
        year = {2026},
        publisher = {Springer Nature Switzerland},
        volume = {LNCS 16893},
        month = {September},
        page = {pending}
}


Reviews

Review #1

  • Please describe the contribution of the paper

    This paper proposes an automated workflow using Implicit Neural Shape Functions (INSF) to regress clinical alignment measurements directly from these latent codes. The proposed method is evaluated on both an internal test dataset and a separate external set from the MRKR dataset to assess its effectiveness.

  • Please list the major strengths of the paper: you should highlight a novel formulation, an original way to use data, demonstration of clinical feasibility, a novel application, a particularly strong evaluation, or anything else that is a strong aspect of this work. Please provide details, for instance, if a method is novel, explain what aspect is novel and why this is interesting.

    1.The proposed method minimizes annotation cost. 2.Blinded repeated measurements by two clinicians from different specialties provide a rigorous human baseline. 3.Decomposing the pipeline into simple, independent stages improves debuggability and suits the limited data regime. 4.Interpretability analysis confirms the latent space encodes physically meaningful anatomical variations.

  • Please list the major weaknesses of the paper. Please provide details: for instance, if you state that a formulation, way of using data, demonstration of clinical feasibility, or application is not novel, then you must provide specific references to prior work.

    1.The method is sensitive to demographic shifts — pediatric skeletal morphology and racial variation are unlikely to be covered by a latent space trained on elderly white TKR patients. 2.The paper did not introduce mechanisms to isolate which stage (SDF prediction, test-time optimization, or angle regression) introduced the error when predictions fail. 3.There is no guarantee that the 128-dimensional latent space is complete for rare or extreme morphologies. 4.The paper does not discuss the regularization design of the latent space or the trade-off between over-constraining and under-constraining the representation. 5.No ablation is provided on latent space design choices, such as dimensionality or regularization strength, and their impact on downstream regression performance. 6.The paper did not report the reconstruction performance based on SDF. Not sure how well the implicit model works.

  • Please rate the clarity and organization of this paper

    Good

  • Please comment on the reproducibility of the paper. Please be aware that providing code and data is a plus, but not a requirement for acceptance.

    The authors claimed to release the source code and/or dataset upon acceptance of the submission.

  • Based on your review and your understanding of the MICCAI Scientific Code of Ethics, do you believe this submission may involve a potential ethics concern or violation?

    N/A

  • Optional: If you have any additional comments to share with the authors, please provide them here. Please also refer to our Reviewer’s guide on what makes a good review and pay specific attention to the different assessment criteria for the different paper categories: https://conferences.miccai.org/2026/en/REVIEWER-GUIDELINES.html

    N/A

  • Rate the paper on a scale of 1-6, 6 being the strongest (6-4: accept; 3-1: reject). Please use the entire range of the distribution. Spreading the score helps create a distribution for decision-making.

    (4) Weak Accept — marginally above the acceptance threshold, but would not mind if rejected, dependent on rebuttal

  • Please justify your recommendation. What were the major factors that led you to your overall score for this paper?

    This paper proposed a practical method for an important clinical problem. The method makes sense, and the result is good.

  • Reviewer confidence

    Somewhat confident (2)

  • [Post rebuttal] After reading the authors’ rebuttal, please state your final opinion of the paper.

    Accept

  • [Post rebuttal] Please justify your final decision from above.

    The rebuttal addressed my concerns.



Review #2

  • Please describe the contribution of the paper

    This paper proposes a deep learning method to estimate automatic knee alignment measurement from 2D radiographs. It regresses the measurement from a compact latent anatomical representation instead of explicit landmark coordinates. Its main significance is conceptual: the framework could extend to measurements defined by overall shape, beyond those based on landmarks. The method is evaluated on a real world dataset against manual measurements. The proposed method achieves performance comparable to state-of-the-art landmark-based approaches, but without a clear accuracy gain. The study is strengthened by meaningful evaluation and honest discussion.

  • Please list the major strengths of the paper: you should highlight a novel formulation, an original way to use data, demonstration of clinical feasibility, a novel application, a particularly strong evaluation, or anything else that is a strong aspect of this work. Please provide details, for instance, if a method is novel, explain what aspect is novel and why this is interesting.

    Extensibility is promising beyond landmark-based measurements. The main strength is the potential generality of the approach. By predicting measurements from latent anatomical embeddings rather than landmarks, the method could be extended to shape-based measurements instead of landmark-based ones, e.g., those relying on lines, planes, or spheres.

    Evaluation is careful and discussion is honest. The experiments are clinically relevant because they compare the method with manual measurements and existing automatic approaches on a real world datasets. Although the method does not outperform the SOTA, it is shown to be competitive. The paper is also strengthened by its clear discussion of results, such as degraded accuracy on external validation data and of why the method does not surpass the SOTA.

  • Please list the major weaknesses of the paper. Please provide details: for instance, if you state that a formulation, way of using data, demonstration of clinical feasibility, or application is not novel, then you must provide specific references to prior work.

    The paper does not show an accuracy gain. The results indicate comparability, not improvement. The method is slightly weaker than reference [12] for most measurements.

    The proposed method appears sensitive to truncated anatomy or image margins. Because manually extending shaft contours significantly improves performance. This raises concerns about robustness in routine clinical images.

    Key design choices are not justified. Some methodological decisions are insufficiently motivated. In particular, the paper replaces latent-code prediction with a U-Net-based SDF prediction followed by latent optimization, but does not explain why this design is preferable. It is unclear whether this choice is principled or simply convenient. Equation 2 is poorly explained: key symbols such as \Omega are undefined, and the relationship between the two summation terms is ambiguous.

    A minor issue is that the compared methods are cited only by reference number, without clearly naming or briefly describing them.

  • Please rate the clarity and organization of this paper

    Poor

  • Please comment on the reproducibility of the paper. Please be aware that providing code and data is a plus, but not a requirement for acceptance.

    The submission does not mention open access to source code or data, but provides a clear and detailed description of the algorithm to ensure reproducibility.

  • Based on your review and your understanding of the MICCAI Scientific Code of Ethics, do you believe this submission may involve a potential ethics concern or violation?

    N/A

  • Optional: If you have any additional comments to share with the authors, please provide them here. Please also refer to our Reviewer’s guide on what makes a good review and pay specific attention to the different assessment criteria for the different paper categories: https://conferences.miccai.org/2026/en/REVIEWER-GUIDELINES.html

    N/A

  • Rate the paper on a scale of 1-6, 6 being the strongest (6-4: accept; 3-1: reject). Please use the entire range of the distribution. Spreading the score helps create a distribution for decision-making.

    (2) Reject — should be rejected, independent of rebuttal

  • Please justify your recommendation. What were the major factors that led you to your overall score for this paper?

    The paper does not show a clear empirical accuracy advantage compared to SOTA. It appears sensitive to truncated anatomy or image margins, limiting practical applicability. Several key designs and equations are not well justified/explained.

  • Reviewer confidence

    Confident but not absolutely certain (3)

  • [Post rebuttal] After reading the authors’ rebuttal, please state your final opinion of the paper.

    Reject

  • [Post rebuttal] Please justify your final decision from above.

    The authors’ responses left most of my concerns unresolved.

    The manuscript didn’t demonstrate performance equivalent to SOTA [12] within an acceptable threshod. SOTA mean absolute difference (MAD), compared with human readings, is about 1.However, the authors specify a threshold of 1 in the equivalence test between MADs of the proposed and SOTA. This implies the proposed method’s error could be up to twice that of SOTA, which appers far from the commont expectation of equivalence.

    Equation 2 appears errorneous. The difference ( S_i - P_{i,k} ), where P_{i,k} is the SDF predition of i-th point of k-th image and S is the groundtruth, should not have one identifcal groundtruth for all different images. The second sum note, \sum_{i=1}^{|X_k|} should be removed. Because this should simply be an objection of reference [21] summed over all samples. The other two sums already iterate both i for points and k for images. The authors’ response miss explaining this issue, indicating a risk of incorrect adaptation of [21] without a sufficient understanding of its underlying meaning.



Review #3

  • Please describe the contribution of the paper

    The authors have proposed an automated workflow based on implicit neural representation for assessing lower-limb alignment for predicting joint health in total knee arthroplasty. While traditional methods are typically manual and time consuming, and machine learning-based approaches are still rigid as well as dependent on manual annotations, the proposed workflow focuses on automation to reduce this dependency on annotations. The authors have validated their workflow against SOTA methods as well as experts (orthopaedic surgeons and radiologists) in the domain.

  • Please list the major strengths of the paper: you should highlight a novel formulation, an original way to use data, demonstration of clinical feasibility, a novel application, a particularly strong evaluation, or anything else that is a strong aspect of this work. Please provide details, for instance, if a method is novel, explain what aspect is novel and why this is interesting.

    The proposed automated workflow reduces the burden on annotators by using implicit neural representations to model the shape and morphology of the bones (femur and tibia) individually without any landmarks. The authors have performed exhaustive experiments to evaluate their workflow against manual inter-rater agreements (orthopaedic surgeons and radiologists) and reported the results with statistical analysis, thereby indicating that the methods perform equivalently. While automated methods have not yet outperformed the manual inter-rater agreement, the proposed workflow demonstrates strong potential as models evolve. They have also claimed to make these clinical measurements for the MRKR dataset publicly accessible which will help the medical community in further research and development.

  • Please list the major weaknesses of the paper. Please provide details: for instance, if you state that a formulation, way of using data, demonstration of clinical feasibility, or application is not novel, then you must provide specific references to prior work.

    While the ideation of implicit neural representations to calculate lower limb angles is novel, the methodology in itself is mostly built on top of existing ideas as cited by the authors. An ablation study would be helpful to understand how the contribution of each component of the proposed framework come together. It would strengthen the paper to justify why a U-Net was trained to predict the SDF and why not any other CNN networks, why was SIREN chosen as the auto-decoder, etc. In Table 1, is there a reason for not including aMPTA and JLCA while evaluating on the internal dataset?

  • Please rate the clarity and organization of this paper

    Satisfactory

  • Please comment on the reproducibility of the paper. Please be aware that providing code and data is a plus, but not a requirement for acceptance.

    The authors claimed to release the source code and/or dataset upon acceptance of the submission.

  • Based on your review and your understanding of the MICCAI Scientific Code of Ethics, do you believe this submission may involve a potential ethics concern or violation?

    N/A

  • Optional: If you have any additional comments to share with the authors, please provide them here. Please also refer to our Reviewer’s guide on what makes a good review and pay specific attention to the different assessment criteria for the different paper categories: https://conferences.miccai.org/2026/en/REVIEWER-GUIDELINES.html

    N/A

  • Rate the paper on a scale of 1-6, 6 being the strongest (6-4: accept; 3-1: reject). Please use the entire range of the distribution. Spreading the score helps create a distribution for decision-making.

    (4) Weak Accept — marginally above the acceptance threshold, but would not mind if rejected, dependent on rebuttal

  • Please justify your recommendation. What were the major factors that led you to your overall score for this paper?

    The authors have proposed an automated workflow to automate the angle calculations and have also exhaustively validated it against manual agreements and compared it with a SOTA method. However, the paper would be strengthened by including ablation studies that give us a proper technical understanding of the workflow.

  • Reviewer confidence

    Confident but not absolutely certain (3)

  • [Post rebuttal] After reading the authors’ rebuttal, please state your final opinion of the paper.

    Accept

  • [Post rebuttal] Please justify your final decision from above.

    The rebuttal has strengthened the paper by adequately addressing the reviewers’ concerns. While I still believe certain aspects of the design could be explained and justified better with experimental evaluation and ablations, the overall contribution is intriguing and relevant to the field.



Author Feedback

We thank the reviewers for their constructive feedback and to resolve misunderstandings, we first want to highlight the key contributions of our work: (i) a multi-structure INR jointly modelling femur and tibia in a shared latent space; (ii) the first landmark-free INR framework for automated lower-limb alignment (LLA) assessment from knee radiographs; (iii) cross-population evaluation on non-US and racially diverse US cohorts; and (iv) release of a novel diverse LLA measurement dataset.

We address specific points below.

Performance (R2): We did not claim improved accuracy; rather, our statistical test shows equivalence between reference and proposed method, as noted by R3.Our contribution is a landmark-free, multi-structure shape representation that reaches the same accuracy ceiling while avoiding dense landmark annotation, extends naturally to non-landmark measurements (lines/contour-based metrics), and generalizes across populations.

Truncated anatomy / shaft extension (R2): We clarify that shaft contours were extended only in SDF labels for UNet training, not in test data. This teaches the model to infer missing shaft regions from truncated radiographs, improving robustness to variable image coverage and generalization to routine clinical images.

Design justification and ablation study (R1,R2,R3): We agree ablations are valuable, but our paper is focused a novel application and validation of INRs rather than on methodology. We use standard components and emphasize clinically meaningful cross-population evaluation. Our experiments showed that SIREN outperformed ReLU MLPs, especially for high-frequency geometric details in curved bone surfaces. UNet was chosen as a simple, robust backbone for spatially coherent SDF prediction [21].

The UNet + test-time optimization decouples image variability from shape estimation: the UNet predicts SDFs and latent optimization fits the shape code. This avoids direct image-to-latent regression, which needs pose/scale/orientation canonicalization and is prone to alignment errors under radiographic variation.

Demographic shifts (R1): The claim that our model fits shapes from “elderly white TKR patients” understates our evaluation. We tested across two countries and multiple racial groups, providing the first cross-racial evaluation of automated LLA assessment. Pediatric LLA is anatomically and clinically distinct, and outside our scope.

Failure isolation (R1): Our pipeline is modular and supports failure diagnosis, as each stage can be inspected independently. Since the method achieved statistical equivalence with manual readers (p < 0.0001 within ±1°), per-stage failure attribution was not pursued.

128-D completeness (R1): The 128-D space balances capacity, accuracy, and interpretability. Individual dimensions strongly correlate with clinical angles (|r|≈0.84 for aTFA; ≈0.79 for JLCA), capturing clinically meaningful variation. Rare-morphology coverage is a general data-driven-model limitation best addressed by including representative training data.

SDF reconstruction (R1): SDF quality is validated downstream: poor SDFs yield inaccurate angles, yet our measurements match expert raters (ICC up to 0.97). Direct SDF reconstruction also achieved a Dice score of 0.96 from overlap between reconstructed and ground-truth contour regions, indicating high accuracy, but was omitted due to space limits.

Eq. 2 / reference clarity (R2): Ω denotes the set of spatial coordinates sampled in the image domain for SDF supervision. The summations are nested: outer over training samples k, inner averaging focal-L1 loss over sampled coordinates. We will further clarify Eq. 2 and references.

Measurement availability (R3): The orthopaedic surgeon measured signed aTFA only, not aMPTA and JLCA on internal data. Due to space limits, we report only internal aTFA, measured by both clinicians. Internal aMPTA and JLCA results for our method were statistically comparable to baselines.




Meta-Review

Meta-review #1

  • Your recommendation

    Invite for Rebuttal

  • Please justify your decision. In case you deviate from the reviewers’ recommendations, explain in detail the reasons why. In case of an invitation for rebuttal, clarify which points are important to address in the rebuttal.

    I recommend Invite for Rebuttal. Reviewer 1 was moderately positive, but highlighted sensitivity to demographic shift and limited analysis of which stage causes failure. Reviewer 2 was clearly negative and argued that the results show comparability rather than improvement, while also questioning the robustness to truncated anatomy and the justification of key design choices. Reviewer 3 was positive overall, but also asked for clearer technical understanding of the workflow. I agree that the proposed implicit shape representation is an interesting idea, but the practical advantage over existing landmark-based methods remains unclear.

  • After you have reviewed the rebuttal and updated reviews, please provide your recommendation based on all reviews and the authors’ rebuttal.

    Accept

  • Please justify your recommendation.

    The paper is not a strong methodological breakthrough and does not clearly improve over SOTA landmark-based methods in accuracy, but it offers a meaningful application contribution, landmark-free lower-limb alignment assessment using a multi-structure implicit shape representation.



Meta-review #2

  • After you have reviewed the rebuttal and updated reviews, please provide your recommendation based on all reviews and the authors’ rebuttal.

    Accept

  • Please justify your recommendation.

    R1 and R3 were both satisfied with the rebuttal and felt that the authors had addressed their concerns. R2, however, remained unconvinced and maintained a Reject recommendation, citing unresolved concerns regarding the equivalence evaluation against SOTA methods and potential issues in the formulation of Equation 2, which the reviewer felt were not adequately clarified in the rebuttal. Both issues should be addressed in the camera-ready version.



Meta-review #3

  • After you have reviewed the rebuttal and updated reviews, please provide your recommendation based on all reviews and the authors’ rebuttal.

    Accept

  • Please justify your recommendation.

    Although one reviewer raised concerns about the accuracy gain over state‑of‑the‑art methods, I find the authors’ rebuttal persuasive, as the proposed method offers a novel landmark‑free, multi‑structure shape representation for measurement. Accordingly, considering the majority of reviewers’ feedback and my own assessment, I am inclined to accept the paper.



back to top